Acid-extractable Phosphorus (9G2)

This test is often referred to as BSES-P. Soil extracts are obtained by shaking milled air-dry soil at a ratio of 1:200 (w/v) with 0.005M sulfuric acid for 16 h. Method 9G1 codes the manual analytical finish of Truog and Meyer, with absorbance at 660 nm. Method 9G2 codes automated analytical finishes (continuous segmented flow and flow injection analysis) based on molybdenum-blue and a preferred absorbance of 882 nm. The methods specify reporting results as mg P/kg on an air-dry basis.
Close

Air dry moisture content (2A1)

This method is required to adjust soil chemical results based on air-dry samples to an oven-dry (105oC)
basis. When the air-dry moisture content (M%) is known, the correction from air-dry
to oven-dry is as follows:

 

            Oven-dry result  =   [Air-dry result  x  (100+ M%)]/100

Close

Nitrate Nitrogen (7C2)

Initially, milled air-dry soil is extracted for 1 h with 2M KCl at a 1:10 soil/solution ratio. For 7C1a to 7C1h methods, mineral-N components are quantified by steam distillations and subsequent titrations. For 7C2 methods, mineral-N fractions in the clarified soil extract are determined by automated colorimetric procedures. Ammonium ions (NH4+) are measured by a modified Berthelot indophenol reaction, while the Griess-Ilosvay reaction is used for NO3-N (and NO2-N). The methods specify reporting results for NH4-N and NO3-N [plus NO2-N if present], respectively, as mg N/kg on an oven-dry (105oC) basis. Specifically, method 7C2a relates to an automated colour, continuous segmented flow analytical finish, while Method 7C2b codes an automated colour finish by flow injection analysis. Little difference is expected in results due to choice of colorimetric finish.

Close

(9.00E+01)

Close

CaCl2 extractable B - ICP (12C2)

The soil is extracted with boiling 0.01M Cacl2 solution at a 1:2 soil:solution ratio for 10 mins.  It is then quickly filtered to avoid any re-fixation of the solubilised B.  The extract is then analysed by ICP-AES.

Close

Electrical conductivity (EC) of 1:5 soil/water extract (3A1)

This test on milled air-dry sample at a soil/water ratio of 1:5 for 1 h is suitable for use on all soils, irrespective of whether acidic or alkaline. It usually underestimates the soluble salt status of soils containing natural or added gypsum, particularly if ³ 1% of gypsum is present. Such soils would have an EC of about 2 dS/m. Soil EC x 0.336 (Method 3B1) approximates percent total soluble salts, while approximate soil ionic strength (Method 3C1) at 0.1 bar (I0.1) can be calculated as follows: I0.1 = [0.0446*EC1:5 – 0.000173], where I0.1 has units of mM, and EC1:5 has units of dS/m @ 25oC.

Close

Exchangeable Acidity (15G1)

 

Exchange acidity (hydrogen and aluminium) by 1M potassium chloride

 
This is a preferred method for estimating the acidic cation status of acidic sub-tropical and tropical soils.

 

Close

Exchangeable Bases - Ammonium Chloride (15A1)

Method for measuring exchangeable bases (Ca2+, Mg2+, Na+, K+) - 1M ammonium chloride at pH 7.0

 

Suited for use on all soils, irrespective of whether acidic or alkaline, but preferred on acidic to weakly alkaline soils not dominated by solid-phase carbonates. Method 15A1 has no pre-treatment to remove soluble salts, with alternatives to remove them chemically (15A2) or to adjust for the presence of soluble sodium (15A3).

Close

Exchangeable Bases - Ammonium Acetate (15D3)

 

Exchangeable bases - 1M ammonium acetate at pH 7.0.

 

This rapid method for exchangeable cations in non-saline acidic through to slightly alkaline soils has no pre-treatment for soluble salts. It should yield similar data to those of method 15A1, except it can overestimate exchangeable Ca in soils containing calcium carbonate.

Close

DTPA Trace Elements (12A1)

Soils are extracted with 0.005M DTPA, at a 1:2 soil:solution ratio, and shaken for 2 hr at 25oC.  Analytical finish is either ICP-AES or Flame AAS.

Close

Colwell extractable K (18A1)

The Colwell extraction was developed for measuring available P, but can also be used for determining available K. The extractant is freshly prepared 0.5M sodium bicarbonate @ pH 8.5, with a soil:extractant ratio of 1:100 and a shaking time of 16 hr.  The sodium cation is considered the primary extractor, displacing the similar potassium cation from exchange sites.

Close

KCl-ext.S (10D1)

Empirical extraction at 40oC in 0.25M KCl, for 3 hours.  Total S in the extract is measured by ICP-AES.

Close

Mehlich3 ICP (18F1)

Analysis of Mehlich 3 extract using inductively-coupled atomic emission spectroscopy (ICP-AES)

Close

Water soluble nitrate - automated colour (7B1)

This method uses the same 1:5 soil/water suspension described for method 3A1. The filtered or centrifuged aliquot is subjected to automated colorimetric analysis based on the Griess-Ilosvay reaction, either by continuous segmented flow analysis (sub-method 7B1a) or by flow injection analysis (sub-method 7B1b) The methods specify reporting nitrate-N (mg N/kg) on an air-dry basis. Note that in some highly weathered soils with a measurable anion exchange capacity, water may not extract all of the adsorbed nitrate-N.
Close

Organic carbon - W & B (6A1)

This measure of soil organic carbon (OCW&B; expressed as %C) usually yields a lower figure than the true total organic carbon value. The method uses finely-milled air-dry sample. It involves wet oxidation by a dichromate-sulphuric acid mixture and relies only on heat of reaction. Soil weight should take account of the expected concentration of OC, and it is expected that allowance will be made for positive soluble Cl‑ interference in soils containing >0.5% Cl. The method specifies reporting on an oven-dry (105oC) basis. Nowadays this method is less preferred than 6B methods.

Close

Phosphorus buffer index, with Colwell P (PBI+ColP) pooled analytical finishes (9I2A)

This index of soil P sorption embraces adsorption as well as precipitation reactions. In addition, it utilises the Colwell-P test on the same soil sample as a measure of current soil P fertility (PBI+ColP). Equilibrium soil extracts are obtained by shaking milled air-dry soil continuously for 17 h at a ratio of 1:10 (w/v) with a P equilibrating solution initially containing the equivalent of 1000 mg P/kg in 0.01M CaCl2. In method 9I2a, orthophosphate-P in final particulate-free supernatant solutions is analysed by a molybdenum-blue analytical finish, at a preferred absorbance of 882 nm. Methods 9I2b and 9I2c code for analytical finishes based on ICPAES and a vanadate-P colour finish, respectively. PBI+ColP is calculated as {[Ps (mg P/kg) + Colwell-P (mg/kg)] / c (mg P/L)0.41}, where Ps = freshly sorbed P and c = final solution P concentration. The methods specify reporting results on an air-dry basis.
Close

PBI - Unadjusted (9I4A)

This index of soil P sorption embraces adsorption as well as precipitation reactions. Equilibrated soil extracts are obtained by shaking milled air-dry soil continuously for 17 h at a ratio of 1:10 (w/v) with a P equilibrating solution initially containing the equivalent of 1000 mg P/kg in 0.01M CaCl2. The residual P in the extract is determined using the Murphy and Riley colorimetric finish.  In other related methods (9I2a, 9I3a) the PBI value is modified to allow for the soil fertility level.  In this method, there is nosuch adjustment. The methods specify reporting results on an air-dry basis.

Close

pH: 1:5 SOIL:0.01M CaCL2 Suspension (4B1)

This pH test on milled air-dry sample is suitable for use on all soils, irrespective of whether acidic or alkaline. Values are usually unaffected by fertilisation prior to sampling, as changes to the soil’s ionic strength is masked by the calcium chloride. Code 4B1 indicates direct use of 0.01M CaCl2, at a soil/solution ratio of 1:5, with mechanical shaking for 1 h prior to pH measurement using calibrated electrodes positioned in the unstirred supernatant after settling of the suspension. Code 4B2 provides a similar measurement outcome but relies on the addition of 0.21M CaCl2 to a 1:5 soil/water suspension to achieve 0.01M CaCl2 prior to measurement of pH as for 4B1. Codes 4B3 and 4B4 are identical to 4B1 and 4B2, respectively, except the soil/CaCl2 suspensions are stirred during measurement. Method 4B5 codes for an MIR surrogate measurement. There is merit in separate use of both water and calcium chloride to measure soil pH.

Close

pH: 1:5 SOIL:0.01M CaCL2 Suspension (4B2)

This pH test on milled air-dry sample is suitable for use on all soils, irrespective of whether acidic or alkaline. Values are usually unaffected by fertilisation prior to sampling, as changes to the soil’s ionic strength is masked by the calcium chloride. Code 4B1 indicates direct use of 0.01M CaCl2, at a soil/solution ratio of 1:5, with mechanical shaking for 1 h prior to pH measurement using calibrated electrodes positioned in the unstirred supernatant after settling of the suspension. Code 4B2 provides a similar measurement outcome but relies on the addition of 0.21M CaCl2 to a 1:5 soil/water suspension to achieve 0.01M CaCl2 prior to measurement of pH as for 4B1.

Codes 4B3 and 4B4 are identical to 4B1 and 4B2, respectively, except the soil/CaCl2 suspensions are stirred during measurement. Method 4B5 codes for an MIR surrogate measurement. There is merit in separate use of both water and calcium chloride to measure soil pH.

Close

pH of 1:5 soil/water suspension (4A1)

This test on milled air-dry sample involves mechanical shaking with deionised water in a closed system for 1 h at a soil/water ratio of 1:5 prior to pH measurement using calibrated electrodes, while stirring the soil/water suspension. The method is suitable for use on all soils, irrespective of whether acidic or alkaline. Values may be lower than expected on recently fertilised soils due to a temporary increase in soil solution ionic strength.
Close

Total organic carbon - high frequency induction furnace, volumetric (6B2)

This method for total soil OC involves production, purification and measurement of CO2 evolved when soil carbon is ignited in a stream of O2. Because all C compounds are converted to CO2, the C from carbonates, charcoal, undecomposed wood, etc, will be included, as no soil pre-treatment is specified. In the volumetric sub-method 6B2a, concentrated KOH solution is used to absorb the CO2 released. The difference between the original volume of gas in the burette and the volume produced after ignition equals the volume of CO2 evolved from the sample, after correction for gas temperature and pressure. Sub-method 6B2b is similar to Method 6B2a, except the CO2 produced by ignition is measured via infrared / thermal conductivity detection. Both 6B2a and 6B2b use finely-milled air-dry sample, with weights varying with expected C concentrations. Surrogate estimates can be obtained by NIR (method 6B4a) or MIR (method 6B4b) reflectance spectroscopy. The methods specify reporting as %C on an oven-dry (105oC) basis.

Close

Total soil N – Dumas high-temperature combustion (7A5)

This method utilises automated and/or microprocessor controlled instrumentation, which mostly is able to measure at least total C, N and S in the same sample. Dumas-N dry oxidation includes all forms of soil N, without the need for lengthy pre-treatments, although results can occasionally differ from those expected in soils with high levels of fixed ammonium-N (lower results) and when soils are organically rich (high results, due to incomplete combustion resulting in the formation of methane rather than CO2). Typically, dry, finely-ground sample is subjected to high-temperature combustion (e.g. 950–1,250oC) in a stream of purified O2. An aliquot of the gases produced by combustion is carried by helium gas to a thermal conductivity cell for measurement of any N2 generated, a process typically taking 3–5 min. A heated copper catalyst reduces NOx to N2. The method specifies reporting results as %N on an oven-dry (105oC) basis.
Close

Total carbon - high frequency induction furnace (with prior physical removal of (6B3)

Following quantitative action / pre-treatment to account for or to physically remove (if present) charcoal and to chemically remove carbonate with excess 5% H2SO3 solution on a hot plate in a fume cabinet, the residual, re-dried soil sample is analysed for soil C by a suitable method, preferably Method 6B2b. The method involving carbonate removal and soil C analysis uses finely-milled air-dry sample, with weights varying with expected C concentrations. The method specifies reporting as %C on an oven-dry (105oC) basis.

Close

Soil organic matter by loss-on-ignition (6G1)

This simple test involves the ignition of finely-milled air-dry sample initially to 105oC and then to 550oC. One hundred times the difference in sample weight in grams between these two temperatures, i.e, [100(Weight105C – Weight550C)] = Loss on ignition550C (LOI550C), which is assumed to approximate % Organic Matter.
Close

Not Specified or mixed (NS)

The laboratory either did not specify the method used or uses a mix of methods when submitting to the certification body

Close

Chloride - 1:5 soil/water extract (5A1)

Tests for water-soluble chloride (Cl) on milled air-dry sample are suitable for use on all soils. For method 5A1, Cl‑ in clarified 1:5 soil/water extracts is determined by potentiometric titration with AgNO3 in conjunction with an Ag/AgNO3 electrode array. For method 5A2a, Cl‑ in clarified 1:5 soil/water extracts is determined by an automated, continuous flow colorimetric procedure based on the formation — in the presence of ferric ions and free thiocyanate ions — of highly coloured ferric thiocyanate in proportion to the Cl‑ concentration. Method 5A2b is similar, except it pertains to the use of flow injection analysis (FIA). For 5A1 and 5A2 methods, it is assumed there are no chemical interferences of significance. Moreover, Method 5A2a has proven more precise than method 5A1, particularly at soil concentrations <50 mg Cl/kg. Other analytical finish options involve chemically-suppressed ion chromatography (5A3a), single-column electronically suppressed ion chromatography (5A3b), and direct measurement by ICPAES (Method 5A4). The methodology specifies reporting results on an air-dry basis.

Close

DE-23

Microwave Digestion - closed system without HF, and final medium HNO3 and/or HCl. ICP-AES finish.

Close

DE-24

Microwave Digestion - closed system without HF, and final medium HNO3 and/or HCl. ICP-MS finish.

Close

CA-37

Finely divided dry sample.  Dumas combustion.

Close

BB-32

Extraction with water.  Titrimetric finish.

Close

BB-31

Extraction with water. Spectrophotometric (automated) finish.