Close

AR Dig. elelements (17C1)

Aqua Regia block digested.  ICP-AES/ICP-MS/GF-AAS finishes.  Produces 'Pseudo-totals'.

Close

Mehlich3 ICP (18F1)

Analysis of Mehlich 3 extract using inductively-coupled atomic emission spectroscopy (ICP-AES)

Close

pH of 1:5 soil/water suspension (4A1)

This test on milled air-dry sample involves mechanical shaking with deionised water in a closed system for 1 h at a soil/water ratio of 1:5 prior to pH measurement using calibrated electrodes, while stirring the soil/water suspension. The method is suitable for use on all soils, irrespective of whether acidic or alkaline. Values may be lower than expected on recently fertilised soils due to a temporary increase in soil solution ionic strength.
Close

Total soil N – Dumas high-temperature combustion (7A5)

This method utilises automated and/or microprocessor controlled instrumentation, which mostly is able to measure at least total C, N and S in the same sample. Dumas-N dry oxidation includes all forms of soil N, without the need for lengthy pre-treatments, although results can occasionally differ from those expected in soils with high levels of fixed ammonium-N (lower results) and when soils are organically rich (high results, due to incomplete combustion resulting in the formation of methane rather than CO2). Typically, dry, finely-ground sample is subjected to high-temperature combustion (e.g. 950–1,250oC) in a stream of purified O2. An aliquot of the gases produced by combustion is carried by helium gas to a thermal conductivity cell for measurement of any N2 generated, a process typically taking 3–5 min. A heated copper catalyst reduces NOx to N2. The method specifies reporting results as %N on an oven-dry (105oC) basis.